Department of Special Education and Communication Disorders
Document Type
Article
Date of this Version
2014
Citation
J Biomech. 2014 July 18; 47(10): 2257–2262.
Abstract
TAC-Cell is a custom-built somatosensory stimulator that delivers pneumatic cutaneous tactile inputs to virtually any skin target on the body and by virtue of its non-ferrous materials is compatible with functional magnetic resonance imaging (fMRI) and magnetoencephalography (MEG) brain scanners. In this study, we describe the method to measure apparent skin displacement induced by TAC-Cell stimulation of the glabrous surface of the distal phalanx of the index finger. Specifically, we studied the effect of four servo controller input voltages (0.4V to 1.0V) on resultant skin displacement among eighteen, neurotypical adult male and female participants. A fiberoptic displacement sensor, commonly used in industrial applications, was coupled to the TAC-Cell to measure the glabrous skin’s kinematic response to different stimulus amplitudes. Skin displacement was significantly dependent on stimulus amplitudes and sex (p< 0.0001). Power spectrum and kinematic analysis of skin displacement showed that the pneumatic TAC-Cell stimulus consists of a spectrally rich, high velocity signal. In related work, we have shown that this dynamic pneumocutaneous stimulus is highly effective in evoking a cortical brain response for neurodiagnostic applications and somatosensory pathway analysis in health and disease.
Comments
© 2014 Elsevier Ltd. All rights reserved.